Know exactly where to measure and trust your findings every time with the FLIR A35sc Benchtop Test Kit. Unlike thermocouples or spot pyrometers, the A35sc allows you to measure thousands of temperature points across your target and get reliable data in seconds. Together with the kit’s lens choices and advanced analysis software, the A35sc kit provides a dependable temperature measurement solution for industrial and R&D labs.
- Products
- Research & Development
- Bench Top Test Kits
- FLIR A35sc Test Kit
Added to cart
-
Fast Data Transfer
A325sc supports 14-bit, 320 × 240 data-streaming over Ethernet at rates as high as 60 Hz.
-
Plug-and-Play
GigE Vision and GenICam compliance streamlines setup, and supports easy integration with third-party software.
-
Advanced Software Options
A325sc works seamlessly with FLIR ResearchIR Max software or MathWorks® Matlab, for intuitive viewing and recording, and advanced processing power.
Contact Sales
Want more information on this product?
Call +1 (866) 477-3687 to speak with a knowledgeable specialist.
Specifications
- IR Resolution
- 320 × 256 pixels
- Resolution
- 320 × 256
- Detector Pitch
- 25 µm
- Spectral Range
- 7.5–13 µm
- Detector Type
- Focal plane array (FPA), uncooled VOX microbolometer
- f-number
- 1.25
- Sensitivity
- <0.05°C @ 30°C/50 mK
- Thermal Sensitivity/NETD
- <0.05°C @ 30°C (86°F) / 50 mK
- Temperature Range
- -25°C to 135°C (-13F to 275°F) / -40°C to 550°C (-40°F to 1022°F)
- Focus
- Fixed
Imaging & Optical
- Detector Pitch
- 25 µm
- Detector Time Constant
- Typical 12 ms
- Detector Type
- Focal plane array (FPA), uncooled VOX microbolometer
- Digital Data Streaming to PC
- Gigabit Ethernet
- Field of view (FOV)
- 48° × 39°
- f-number
- 1.25
- Focal Length
- 9 mm (0.35 in.)
- Focus
- Fixed
- Image Frequency
- 60 Hz
- IR Resolution
- 320 × 256 pixels
- Minimum Focus Distance
- 3.2 cm (1.6 in.)
- Minimum Spot Size
- 100 µm
- Resolution
- 320 × 256
- Sensitivity
- <0.05°C @ 30°C/50 mK
- Spatial resolution (IFOV)
- 2.78 mrad
- Spectral Range
- 7.5–13 µm
- Thermal Sensitivity/NETD
- <0.05°C @ 30°C (86°F) / 50 mK
Measurement & Analysis
- Accuracy
- ±5°C (±9°F) or ±5% of reading
- Atmospheric transmission correction
- Automatic, based on inputs for distance, atmospheric temperature and relative humidity
- Camera size (L x W x H)
- 106 × 40 × 43 mm (4.2 × 1.6 × 1.7 in.)
- Emissivity Correction
- Variable from 0.5 to 1.0
- External Optics & Windows Correction
- Automatic, based on input of optics/window transmission and temperature
- Measurement Corrections
- Global object parameters
- Object Temperature Range
- -25°C to 135°C (-13 to 275°F) / –40°C to 550°C (-40 to 1022°F)
- Optics Transmission Correction
- Automatic, based on signals from internal sensors
- Reflected apparent temperature correction
- Automatic, based on input of reflected temperature
- Temperature Range
- -25°C to 135°C (-13F to 275°F) / -40°C to 550°C (-40°F to 1022°F)
- Time vs Temperature Plotting
- Yes
Communication & Data Storage
- Digital I/O connector type
- 12-pole M12 connector (shared with Digital Synchronization and External power)
- Digital I/O isolation voltage
- 500 VRMS
- Digital I/O supply voltage
- 2–40 VDC, max 200 mA
- Digital input purpose
- General purpose
- Digital Inputs
- General Purpose, 1× opto-isolated, "0" <1.2 VDC, "1" = 2–25 VDC.
- Digital Output Purpose
- General purpose output to ext. device (programmatically set)
- Digital Outputs
- 1× opto-isolated, 2–40 VDC, max 185 mA
- Digital Synchronization Connector Type
- 12-pole M12 connector (shared with Digital I/O and External power)
- Ethernet
- Control and image
- Ethernet Communication
- GigE Vision ver. 1.2 Client API GenICam compliant
- Ethernet Connector Type
- RJ-45
- Ethernet Image Streaming
- 8-bit monochrome @ 60 Hz •Signal linear/ DDE •Automatic/ Manual •Flip H&V 14-bit 320 × 256 pixels @ 60 Hz •Signal linear/ DDE •Temperature linear GigE Vision and GenICam compatible
- Ethernet Power
- Power over Ethernet, PoE IEEE 802.3af class 0 Power
- Ethernet Protocols
- TCP, UDP, ICMP, IGMP, DHCP, GigEVision
- Ethernet Standard
- IEEE 802.3
- Ethernet Type
- Gigabit Ethernet
- Synchronization in
- 1×, non-isolated
- Synchronization in - purpose
- Frame sync in to control camera
- Synchronization in - type
- LVC Buffer @3.3V, "0" <0.8 V, "1" >2.0 V.
- Synchronization out
- 1×, non-isolated
- Synchronization out - purpose
- Frame sync out to control another Ax5 camera
- Synchronization Out - type
- LVC Buffer @ 3.3V, "0"=24 MA max, "1"= –24 mA max.
General
- Base Mounting
- 4 × M3 thread mounting holes (bottom)
- Housing material
- Magnesium and aluminum
- Package Includes
- Hard transport case Infrared camera with lens Base support Cable tie (2 ea.) Ethernet cable CAT-6, 2m/6.6 ft (2 ea.) FLIR ResearchIR Standard 4 Focus adjustment tool Gooseneck Mains cable kit (UK,EU,US) PoE Injector (power over Ethernet) Printed documentation Table stand
- Packaging Size
- 370 x 180 x 320 mm (14.6 x 7.1 x 12.6 in.)
- Packaging Type
- Cardboard box
- Research IR Software Included
- Yes
- Weight
- 0.200 kg (0.44 lb.)
Power
- External Power Connector Type
- 12-pole M12 connector (shared with Digital I/O and Digital Synchronization )
- External Power Operation
- 12/24 VDC, <3.5 W nominal <6.0 W absolute max
- Voltage
- Allowed range 10–30 VDC
Environmental & Certifications
- EMC
- •EN 61000-6-2 (Immunity) •EN 61000-6-3 (Emission) •FCC 47 CFR Part 15 Class B (Emission)
- Encapsulation
- IP 40 (IEC 60529) with base support mounted
- Humidity (Operating and Storage)
- IEC 60068-2-30/24 h 95% relative humidity 25°C to 40°C (77°F to 104°F)
- Operating Temperature Range
- –15°C to 50°C (5°F to 122°F) The operating temperature range assumes that the camera is mounted on the base support (included in the package) or a similar type of heatsink.
- Shock
- 25 g (IEC 60068-2-27)
- Storage Temperature Range
- –40°C to 70°C (–40°F to 158°F)
- Tripod Mounting
- UNC ¼"-20 (on three sides)
- Vibration
- 2 g (IEC 60068-2-6 & MIL-STD810G)
Find a Dealer
Find a Dealer
Ready to buy? See our list of dealers.
Export Restrictions
Export Restrictions
The information contained in this page pertains to products that may be subject to the International Traffic in Arms Regulations (ITAR) (22 C.F.R. Sections 120-130) or the Export Administration Regulations (EAR) (15 C.F.R. Sections 730-774) depending upon specifications for the final product; jurisdiction and classification will be provided upon request.
FLIR A35sc Test Kit
Want more information on this product?
Call +1 (866) 477-3687 to speak with a knowledgeable specialist.